Manually Triggered Ventilator (D-MTV)

Manually Triggered Ventilator (D-MTV)

611-3101

Used to provide high-concentration oxygen for unconscious, non-breathing and for conscious, breathing injured divers

 PRODUCT DETAILS

  • Suitable for use with all DAN oxygen units
  • Flow-restricted, oxygen-powered ventilator provides oxygen as both a demand valve and for a non-breathing, unconscious diver.
  • MTV valve hose connection is a quick connect end fitting, making it more flexible and quicker to assemble and disassemble.
  • Supplied with 6 ft hose, with a DISS connector on one end to connect to a standard multifunction regulator equipped for demand valve supply.,
  • Contains built-in safety features such as over-pressure relief.
  • Hard-wearing, robust thermoplastic body with stainless steel internal parts.
  • Robust to minimized damage from handling
  • Corrosion free
  • Superior performance with reduced breathing resistance
  • No periodic maintenance requirements

SPECIFICATIONS

  • Required operating pressure: 40 – 80 psi (2.7 – 5.5 bar)
  • Maximum continuous flow capacity 165 LPM (peak 200 LPM) at 65 psi (4.5 bar)
  • Cracking (trigger) pressure: < 1 cm H2O (< 0.1 kPa or 0.7 mmHg)
  • Exhalation resistance: < 3 cm H2O (< 0.3 kPa or 2.2 mmHg)
  • Normal by-pass pressure set to ± 20 cmH2O (< 2.0 kPa or 14.7 mmHg) to protect lungs from potential barotrauma if trigger not released
  • Limiting pressure: < 60 cm H2O (< 6 kPa or 44 mmHg) through a secondary safety device
  • Inlet filter: 25 microns

Note that these MTV valves are not compatible with the old hoses.